(diplomsko delo)
Aleš Avguštin (Author), Miljenko Križmarić (Mentor), Mateja Lorber (Co-mentor)

Abstract

V diplomskem delu smo izvedli validacijo simulatorja DIALOG plus glede na njegove funkcije, kjer smo simulacijo primerjali z realnim dializnim aparatom. Ugotavljamo, da se vizualni del simulacije popolnoma ujema z realnim aparatom. Simulator ne dosega in izpolnjuje vseh faz izobraževanja, ki so jih določile medicinske sestre za uvajanje v delo novozaposlenih na oddelku za hemodializo. Ugotovili smo, da s pomočjo teh simulacij ne pridobimo dovolj znanja in izkušenj, katere bi bile potrebne za samostojno delo z realnim aparatom. S simulatorjem lahko posamezniku predstavimo pomen oznak na krmilnem delu aparata in njihovih vrednosti, ter jih z grafičnim prikazom poučimo o hemodializnem postopku. Glavna prednost simulatorja je v tem, da kljub napakam ne moremo škodovati bolniku.

Keywords

hemodializa;simulacije;dializni aparat;izobraževanje;

Data

Language: Slovenian
Year of publishing:
Source: Maribor
Typology: 2.11 - Undergraduate Thesis
Organization: UM MF - Faculty of Medicine
Publisher: [A. Avguštin]
UDC: 616.6
COBISS: 1854116 Link will open in a new window
Views: 1759
Downloads: 330
Average score: 0 (0 votes)
Metadata: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Other data

Secondary language: English
Secondary title: VALIDATION OF THE DIALOG PLUS DIALYSIS MACHINE SIMULATION
Secondary abstract: In this thesis we performed validation of the simulator DIALOG plus in relation to its functions, where we compare the simulation with real dialysis machine. We find that the visual part of the simulation matches perfectly with the real machine. The simulator does not reach and fulfill all phases of education which were identified by nurses for the induction of new employees in the department of hemodialysis. We found that with the help of these simulations we not obtained sufficient knowledge and experience, which would be required to work independently with real machine. With the simulator we presented to an individual the meaning of markings on the control of the device, and their values and by the graphic learn about hemodialysis procedure. The main advantage of a simulator is that despite the mistakes we can not harm the patient.
Secondary keywords: hemodialysis;simulations;dialysis machine;education;
URN: URN:SI:UM:
Type (COBISS): Undergraduate thesis
Thesis comment: Univ. v Mariboru, Fak. za zdravstvene vede
Pages: IV, 61 f., 7 f. pril.
Keywords (UDC): applied sciences;medicine;technology;uporabne znanosti;medicina;tehnika;medical sciences;medicina;pathology;clinical medicine;patologija;klinična medicina;pathology of the urogenital system;urinary and sexual (genital) complaints;urology;patologija urogenitalnega sistema;bolezni sečil in spolnih organov;urologija;
ID: 15931
Recommended works:
, no subtitle data available
, who will get it and how?