Abstract
We present the applicability of a new ultra sensitive analytical method for the simultaneous determination of biliverdin and bilirubin in human serum. The method comprises isocratic reversed-phase(RP) C18 high-performance liquid chromatography (HPLC) and thermal lens spectrometric detection (TLS)based on excitation by a krypton laser emission line at 407 nm. This method enables the separation of IX-α biliverdin and IX-α bilirubin in 11 min. with limit of detection (LOD) and limit of quantitation (LOQ) for
biliverdin of 1.2 nM and 3 nM, and 1nM and 2.8 nM for bilirubin,respectively.In addition, a step-gradient elution was set up, by changing the mobile phase composition, in order to further enhance the sensitivity for bilirubin determination with LOD and LOQ of 0.5 nM and 1.5 nM, respectively. In parallel, an isocraticHPLC-DAD method was developed for benchmarking against HPLC-TLS methods. The LOD
and LOQ forbiliverdin were 6 nM and 18 nM, and 2.5 nM and 8nM for bilirubin,respectively. Additionally, both isocratic methods were applied for measuring biliverdin and free bilirubin in human serum samples (from 2 male and 2 female healthy donors). Combining isocratic HPLC method with TLS
detector was crucial for first ever biliverdin determination in serum together with simultaneous free bilirubin determination. We showed for the first time the concentration ratio of free bilirubin versus unbound biliverdin in human serum samples.
Keywords
bilirubin;HPLC;TLS;biliverdin;serum;
Data
Language: |
English |
Year of publishing: |
2016 |
Typology: |
1.01 - Original Scientific Article |
Organization: |
UNG - University of Nova Gorica |
UDC: |
543.2/.9 |
COBISS: |
4178683
|
ISSN: |
0039-9140 |
Views: |
4774 |
Downloads: |
0 |
Average score: |
0 (0 votes) |
Metadata: |
|
Other data
URN: |
URN:SI:UNG |
Type (COBISS): |
Not categorized |
Pages: |
str. 92-98 |
Issue: |
ǂVol. ǂ154 |
Chronology: |
Jul. 2016 |
DOI: |
10.1016/j.talanta.2016.03.053 |
ID: |
9132433 |