Miha Hiti (Author)

Abstract

This paper describes a validation of a combinatorial calibration technique based calibration procedure for strain-gauge amplifier calibration. The deviation between two strain-gauge amplifiers at calibration using the combinatorial technique is compared to the deviation between the same two amplifiers at calibration on a force calibration machine as a transducer-amplifier chain under unchanged conditions, serving as an amplifier linearity comparator. This enables the validation to be confirmed with a suitable expanded measurement uncertainty of 3 nV/V.

Keywords

ojačevalniki za uporovne lističe;kalibriranje;kombinatorno kalibriranje;validacija;strain-gauge amplifier;calibration;combinatorial technique;validation;

Data

Language: English
Year of publishing:
Typology: 1.01 - Original Scientific Article
Organization: ZAG - Slovenian National Building and Civil Engineering Institute
Publisher: International Measurement Confederation
UDC: 621.317
COBISS: 54802947 Link will open in a new window
ISSN: 2221-870X
Views: 15
Downloads: 4
Average score: 0 (0 votes)
Metadata: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Other data

Secondary language: Slovenian
Secondary keywords: ojačevalniki za uporovne lističe;kalibriranje;kombinatorno kalibriranje;validacija;
Source comment: Opis vira z dne: 11. 3. 2021; Nasl. z nasl. zaslona;
Pages: str. 205-209
Volume: ǂVol. ǂ9
Issue: ǂno. ǂ5
Chronology: Dec. 2020
DOI: 10.21014/acta_imeko.v9i5.970
ID: 19829964