Miha Hiti (Avtor)

Povzetek

This paper describes a validation of a combinatorial calibration technique based calibration procedure for strain-gauge amplifier calibration. The deviation between two strain-gauge amplifiers at calibration using the combinatorial technique is compared to the deviation between the same two amplifiers at calibration on a force calibration machine as a transducer-amplifier chain under unchanged conditions, serving as an amplifier linearity comparator. This enables the validation to be confirmed with a suitable expanded measurement uncertainty of 3 nV/V.

Ključne besede

ojačevalniki za uporovne lističe;kalibriranje;kombinatorno kalibriranje;validacija;strain-gauge amplifier;calibration;combinatorial technique;validation;

Podatki

Jezik: Angleški jezik
Leto izida:
Tipologija: 1.01 - Izvirni znanstveni članek
Organizacija: ZAG - Zavod za gradbeništvo Slovenije
Založnik: International Measurement Confederation
UDK: 621.317
COBISS: 54802947 Povezava se bo odprla v novem oknu
ISSN: 2221-870X
Št. ogledov: 15
Št. prenosov: 4
Ocena: 0 (0 glasov)
Metapodatki: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Ostali podatki

Sekundarni jezik: Slovenski jezik
Sekundarne ključne besede: ojačevalniki za uporovne lističe;kalibriranje;kombinatorno kalibriranje;validacija;
Komentar vira: Opis vira z dne: 11. 3. 2021; Nasl. z nasl. zaslona;
Strani: str. 205-209
Letnik: ǂVol. ǂ9
Zvezek: ǂno. ǂ5
Čas izdaje: Dec. 2020
DOI: 10.21014/acta_imeko.v9i5.970