diplomsko delo
Povzetek
Uvod: Rentgensko slikanje je najpogosteje uporabljena medicinska metoda slikanja.
Slikanje prsnih organov je ena najpogostejših preiskav v splošni radiografiji. Zaradi pogoste
izvedbe je tudi ena najbolj dozno obremenjujočih. Posebej pogosti postopki so rentgenski
posnetki prsnega koša v postelji, kjer se slikanje izvaja v ležečem položaju s slikovnim
sprejemnikom neposredno pod pacientom. Namen: Namen diplomskega dela je ugotoviti
razlike v doznih obremenitvah (DAP, efektivna doza, doza na organe, EI) in kakovosti slike
pri slikanju prsnih organov leže, z uporabo različnih ekspozicijskih pogojev ter
spreminjanjem položaja slikovnega sprejemnika. Metode dela: Diplomsko delo smo
naredili s pregledom obstoječe literature in eksperimentalnim delom. Eksperimentalni del
smo izvedli s pomočjo antropomorfnega fantoma PBU 60 in rentgenskim aparatom
Multix/Vertix proizvajalca Siemens, na Zdravstveni fakulteti. Uporabili smo DR in CR
sistema. Opravili smo 210 slikanj prsnih organov, pri katerih smo uporabili ekspozicijske
pogoje od 81 do 90 kV ter od 0,50 do 1,0 mAs. RGO je bila pri vseh položajih SS 82 cm.
Vstop centralnega žarka je potekal v anteroposteriorni smeri. Pri posameznih meritvah smo
opazovali DAP vrednost ter ekspozicijski indeks. Kasneje smo izračunali še efektivno dozo,
SNR ter CNR vrednost. Rezultati: Na podlagi rezultatov smo ugotovili, da tako kot pri DR,
kot tudi pri CR sistemu obstajajo statistično značilne razlike v ekspozicijskem indeksu. Pri
DR sistemu obstajajo statistične razlike med P1 (položaj slikovnega sprejemnika pod
fantomom) in P3 (položaj slikovnega sprejemnika pod mizo) (p < 0,001). Razlika v
povprečju je znašala 100,15 (48,77 %). Pri CR sistemu pa smo ugotovili, da obstajajo
statistično značilne razlike pri vseh položajih. Največja razlika v povprečju je bila med P1
in P3, 138,59 (64,77 %). Pri DAP vrednosti in efektivni dozi ni bilo statistično značilnih
razlik med DR in CR sistemom. Za vrednosti CNR pri DR sistemu so statistično značilne
razlike med P1 in P2 (položaj slikovnega sprejemnika pod podlogo) ter med P1 in P3 .
Največja razlika v povprečju je med P1 in P2 4,01 dB (12,96 %). Pri CR sistemu so
statistično značilne razlike med P1 in P3 ter med P2 in P3. Največja razlika v povprečju je
med P2 in P3 1,28 dB (4,22 %). Pri DR sistemu obstajajo statistično značilne razlike v SNR
vrednostih med P1 in P2 ter med P1 in P3. Največja razlika v povprečju je med P1 in P2 4,6
dB (13,91 %). Pri CR sistemu, pa obstajajo statistično značilne razlike med P2 in P3 ter med
P1 in P3. Največja razlika v povprečju je bila med P1 in P3 0,99 dB (3,04 %). Razprava in
zaključek: Dokazali smo, da na vrednost ekspozicijskega indeksa ter na kakovost slike
vpliva spremenjen položaj slikovnega sprejemnika in spremenjeni ekspozicijski pogoji,
medtem ko na DAP vrednost položaj slikovnega sprejemnika nima vpliva. Na DAP vrednost
smo vplivali le s spremembo ekspozicijskih pogojev.
Ključne besede
diplomska dela;radiološka tehnologija;prsni organi;rentgensko slikanje;AP;doza;
Podatki
Jezik: |
Slovenski jezik |
Leto izida: |
2022 |
Tipologija: |
2.11 - Diplomsko delo |
Organizacija: |
UL ZF - Zdravstvena fakulteta |
Založnik: |
[L. Hankić |
UDK: |
616-07 |
COBISS: |
121520131
|
Št. ogledov: |
12 |
Št. prenosov: |
4 |
Ocena: |
0 (0 glasov) |
Metapodatki: |
|
Ostali podatki
Sekundarni jezik: |
Angleški jezik |
Sekundarni naslov: |
Supine chest radiography |
Sekundarni povzetek: |
Introduction: X-ray imaging is the most commonly used medical imaging method. Chest
imaging is one of the most common examinations in general radiography. Due to its frequent
performances it is also one of the most dose-loading. Particularly common procedures are
chest X-rays in bed, where imaging is performed in a supine position with an image detector
directly below the patient. Purpose: The purpose of this thesis was to determine the
differences in dose loads (DAP, effective dose, dose to organs, EI) and image quality when
imaging the thoracic organs lying down, using different exposure conditions and changing
the position of the image detector. Methods: We did our diploma work with a review of
existing literature and experimental work. The experimental part was performed on a
anthropomorphic phantom PBU 60 and a Multix / Vertix X-ray machine manufactured by
Siemens, at the Faculty of Medicine. We used DR and CR systems. We performed 210 chest
organ imaging, using exposure conditions ranging from 81 to 90 Kv and from 0.50 to 1.0
mAs. The RGO was 82 cm in all SS positions. The phantom was placed in the AP projection.
The DAP value and the exposure index were observed for individual measurements. Later,
trough programs, we also calculated the effective dose, SNR and CNR value. Results:
Based on the results, we found that both in the DR and in the CR system, there are
statistically significant differences in the exposure index. In the DR system, there are
statistical differences between P1 (position of the image receiver under the phantom) and P3
(position of the image receiver under the table) (p <0.001). the difference averaged 100.15
(48.77%). In the CR system, however, we found that there were statistically significant
differences in all positions. The largest difference on average was between P1 and P3, 138.59
(64.77%). There were no significant statistical differences between the DAP value and the
effective dose in the DR and CR systems. Differences between P1 and P2 and between P1
and P3 are statistically significant for the CNR values of the DR system. The largest average
difference between P1 and P2 (position of the image receiver under the lining) is 4.01 dB
(12.96%). With the CR system, there are statististically significant differences between P1
and P3 and between P2 and P3. The largest average difference between P2 and P3 is 1.28
Db (4.22%). With the DR system, there are statistically significant differences in SNR values
between P1 and P2 and between P1 and P3. The largest average difference between P1 and
P2 is 4.6 Db (13.91%). In the case of the CR system, there are statistically significant
differences between P2 and P3 and between P1 and P3. The largest average difference was
between P1 and P3 of 0.99 dB (3.04%). Discussion and conclusion: We proved that the
value of the exposure index and the image quality are affected by the changed position of
the image receiver and the changed exposure conditions, while the DAP value is not affected
by the position of the image receiver. The DAP value was influenced only by changing the
exposure conditions. |
Sekundarne ključne besede: |
diploma theses;radiologic technology;thoracic organs;X-ray;AP;dose; |
Vrsta dela (COBISS): |
Diplomsko delo/naloga |
Študijski program: |
0 |
Komentar na gradivo: |
Univ. v Ljubljani, Zdravstvena fak., Oddelek za radiološko tehnologijo |
Strani: |
36 str. |
ID: |
16458534 |