frequency and causes during routine testing
Cleo Keppens (Avtor), Yann Van Royen (Avtor), Anne Brysse (Avtor), Sophie Cotteret (Avtor), Estrid Høgdall (Avtor), Tine Plato Kuhlmann (Avtor), Brendan O'Sullivan (Avtor), Patrick Pauwels (Avtor), Siegrid Pauwels (Avtor), Mitja Rot (Avtor)

Povzetek

Context. Errors in laboratory medicine could compromise patient safety. Good laboratory practice includes identifying and managing nonconformities in the total test process. Varying error percentages have been described in other fields but are lacking for molecular oncology. Objectives. To gain insight into incident causes and frequency in the total test process from 8 European institutes routinely performing biomarker tests in non-small cell lung cancer and colorectal cancer. Design. All incidents documented in 2018 were collected from all hospital services for pre-preanalytical entries before the biomarker test, as well as specific incidents for biomarker tests. Results. There were 5185 incidents collected, of which 4363 (84.1%) occurred in the pre-preanalytical phase (all hospital services), 2796 of 4363 (64.1%) related to missing or incorrect request form information. From the other 822 specific incidents, 166 (20.2%) were recorded in the preanalytical phase, 275 (33.5%) in the analytical phase, and 194 (23.6%) in the postanalytical phase, mainly due to incorrect report content. Only 47 of 822 (5.7%) incidents were recorded in the post-postanalytical phase, and 123 (15.0%) in the complete total test process. For 17 of 822 (2.1%) incidents the time point was unknown. Prepreanalytical incidents were resolved sooner than incidents on the complete process (mean 6 versus 60 days). For 1215 of 5168 (23.5%) incidents with known causes a specific action was undertaken besides documenting them, not limited to accredited institutes. Conclusions. There was a large variety in the number and extent of documented incidents. Correct and complete information on the request forms and final reports are highly error prone and require additional focus.

Ključne besede

molekularna patologija;molecular pathology;

Podatki

Jezik: Angleški jezik
Leto izida:
Tipologija: 1.01 - Izvirni znanstveni članek
Založnik: College of American Pathologists
UDK: 616-07
COBISS: 54257411 Povezava se bo odprla v novem oknu
ISSN: 1543-2165
Št. ogledov: 591
Št. prenosov: 145
Ocena: 0 (0 glasov)
Metapodatki: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Ostali podatki

Sekundarne ključne besede: Pathology;Laboratories, hospital;Laboratory techniques and procedures;Diagnostic errors;Patologija;Bolnišnični laboratoriji;Laboratorijske tehnike in postopki;Diagnostične napake;
Komentar vira: Nasl. z nasl. zaslona; Soavtor iz Slovenije: Mitja Rot; Opis vira z dne 8. 3. 2021;
Strani: str. 1270–1279
Letnik: ǂVol. ǂ145
Zvezek: ǂiss. ǂ10
Čas izdaje: Oct. 2021
DOI: 10.5858/arpa.2020-0152-OA
ID: 12632286
Priporočena dela:
, frequency and causes during routine testing
, methods and protocols