diplomsko delo
Povzetek
Uvod: Radioterapija predstavlja pomembno vlogo pri zdravljenju bolnikov z rakom glave
in vratu. Med obsevanjem se lahko pri bolnikih pojavijo anatomske spremembe, ki lahko
privedejo do premajhnega odmerka doze na tarčni volumen in prevelikega odmerka doze na
kritične organe. V primeru prevelikih odstopanj je pomembno, da se naredi ponovna
priprava bolnika na obsevanje na računalniško tomografskem (CT) simulatorju. Namen: Z
raziskavo smo želeli ugotoviti, koliko je bilo ponovnih priprav pri bolnikih z rakom glave in
vratu na aparatu Halcyon 3.0 (Varian), zakaj je do njih prišlo in pri kateri frakciji so se
pojavile. Prav tako je namen preveriti, ali z uporabo geometrične verifikacije s kilovoltno
(kV) računalniško tomografijo s stožčastim snopom (CBCT) odkrijemo več primerov za
ponovno pripravo kot na Unique Performance Edition (Varian), kjer se uporablja MV
slikovna verifikacija z uporabo sistema za elektronsko portalno slikanje (EPID). Metode
dela: Retrospektivno raziskavo smo izvedli na Onkološkem Inštitutu v Ljubljani na oddelku
za teleradioterapijo. Pregledali smo podatke bolnikov z rakom glave in vratu, ki so bili od
novembra 2020 do marca 2022 obsevani na obsevalnem aparatu Halcyon 3.0 (Varian). Na
podlagi podatkov smo raziskali, koliko je bilo ponovnih priprav na obsevanje, pri kateri
frakciji so se pojavile in zakaj je do njih prišlo ter izvedli primerjavo med dvema različnima
geometričnima verifikacijama na obeh aparatih. Rezultati: Od 203 bolnikov na Halcyon 3.0
(Varian) je bila pri 30 bolnikih (15%) izvedena ponovna priprava na obsevanje na CT
simulatorju. Rezultati so pokazali, da je najpogostejši vzrok za to izguba telesne teže med
obsevanjem (30 %). Ponovna priprava je bila najpogosteje izvedena že pred/po 1. frakciji in
pri 2.-5. frakciji zaradi neujemanja položaja hrbtenice, zmanjšanja tumorja ali pa bolnik ni
mogel zdržati pod fiksacijsko masko. Razprava in zaključek: Anatomske spremembe, ki
se pojavijo med obsevanjem bolnikov z rakom glave in vratu, so najpogosteje posledica
izgube telesne teže. Za prepoznavo anatomskih sprememb in zagotavljanja natančnega
obsevanja se je geometrična verifikacija kV CBCT izkazala za bolj učinkovito v primerjavi
z MV EPID sistemom. Za nadaljnjo raziskovanje bi bilo priporočljivo izvesti še primerjavo
med obema geometričnima verifikacijama za različne vrste rakov
Ključne besede
diplomska dela;radiološka tehnologija;radioterapija;verifikacija;obsevalne tehnike;rak glave in vratu;ponovna priprava;anatomske spremembe;
Podatki
Jezik: |
Slovenski jezik |
Leto izida: |
2022 |
Tipologija: |
2.11 - Diplomsko delo |
Organizacija: |
UL ZF - Zdravstvena fakulteta |
Založnik: |
[L. Pogorevčnik |
UDK: |
616-07 |
COBISS: |
120487171
|
Št. ogledov: |
36 |
Št. prenosov: |
15 |
Ocena: |
0 (0 glasov) |
Metapodatki: |
|
Ostali podatki
Sekundarni jezik: |
Angleški jezik |
Sekundarni naslov: |
Analysis of repeated patient CT simulations for head and neck cancer on linear accelerator Halcyon 3.0 (Varian) |
Sekundarni povzetek: |
Introduction: Radiotherapy plays an important role in the treatment of patients with head
and neck cancer. During irradiation, some anatomical changes can occur leading to
underdose on the target and overdose on critical organs. In case of excessive deviations, it is
important to make repeated patient computed tomography (CT) simulations. Purpose: The
purpose of the diploma work was to analyze, how many repeated patient CT simulations for
head and neck cancer on linear accelerator Halcyon 3.0 (Varian) there were, why they
occurred, and in which fraction they appeared. We also wanted to research, whether we can
discover more repeated patient CT simulations using geometric verification with kilovolt
(kV) cone beam CT (CBCT) or with MV imaging verification using electronic portal
imaging (EPID) on Unique Performance Edition (Varian). Methods: Retrospective research
was conducted at the Institute of Oncology Ljubljana at the Department of Teleradiotherapy.
We reviewed the data of patients with head and neck cancer from November 2020 till March
2022, who were irradiated on Halcyon 3.0 (Varian). Based on the data, we performed an
analysis of how many repeated patient CT simulations there were, why they occurred, and
in which fraction they appeared and made a comparison between two different geometric
verifications on each of the linear accelerators. Results: Of the 203 patients on Halcyon 3.0
(Varian), 30 patients (15%) underwent repeated patient CT simulations. The results showed
that the most common reason for that was weight loss during irradiation (30%). Repeated
patient CT simulations were most often performed before / after the 1st fraction and in the
2nd-5th fraction due to mismatch of the spine position, tumor reduction or the patient could
not stand the fixation mask. Discussion and conclusion: Anatomical changes, that occur
during irradiation of patients with head and neck cancer, most often happen due to weight
loss. To identify anatomical changes and ensure accurate irradiation, geometric verification
of kV CBCT has proven to be more effective compared to the MV EPID system. For further
research, it would be recommended to make a comparison between the two geometric
verifications for different types of cancers. |
Sekundarne ključne besede: |
diploma theses;radiologic technology;radiotherapy;verification;radiation techniques;head and neck cancer;repeated patient CT simulations;anatomical changes; |
Vrsta dela (COBISS): |
Diplomsko delo/naloga |
Študijski program: |
0 |
Komentar na gradivo: |
Univ. v Ljubljani, Zdravstvena fak., Oddelek za radiološko tehnologijo |
Strani: |
32 str., [3] str. pril. |
ID: |
16345584 |