magistrsko delo
Daša Lučić (Avtor), Damijan Škrk (Recenzent), Nejc Mekiš (Mentor)

Povzetek

Uvod: Klasično rentgensko slikanje vratne hrbtenice je primarna metoda za diagnosticiranje poškodb vratne hrbtenice. Kakovost rentgenograma je odvisna od različnih dejavnikov, kot so izkušnje radiološkega inženirja, značilnosti opreme in tehnika slikanja. Ena izmed metod za izboljšanje kakovosti slike je uporaba radiografske rešetke, ki zmanjša sipano sevanje in poveča kontrast, a hkrati poveča dozo sevanja za pacienta. Namen: Cilj raziskave je bil ugotoviti vpliv odstranitve radiografske rešetke pri stranskem slikanju vratne hrbtenice na dozo sevanja in kakovost rentgenograma. Metode dela: Raziskava je bila izvedena v dveh delih – prvi del je vključeval meritve na fantomu, drugi del pa meritve na pacientih v kliničnem okolju. Uporabljene so bile deskriptivna metoda in prospektivna študija, pri čemer so za oceno kakovosti rentgenogramov bili uporabljeni standardizirani kriteriji, meritev doze pa je bila izvedena z DAP metrom. Rezultati: Rezultati so pokazali, da odstranitev rešetke pri slikanju vratne hrbtenice vodi do statistično značilnega (p < 0,001) znatnega zmanjšanja doze sevanja (63 % manjša doza), pri čemer ni bilo ugotovljenih pomembnih razlik v kakovosti rentgenogramov (p = 0,192). Statistično pomembnih razlik v prikazu anatomskih struktur ni bilo ugotovljenih (p = 0,226). Vizualizacija anatomskih struktur je ostala primerljiva, kar kaže, da je mogoče ohraniti diagnostično uporabno kakovost slike tudi brez uporabe rešetke. Razprava in zaključek: Glede na rezultate lahko sklepamo, da je slikanje brez rešetke učinkovita alternativa, ki omogoča optimizacijo radioloških protokolov z zmanjšanjem doze za pacienta, ne da bi pri tem poslabšali kakovost slik. Vpeljava takšnega pristopa bi lahko pomembno prispevala k izboljšanju varnosti pacientov pri radioloških preiskavah.

Ključne besede

magistrska dela;radiološka tehnologija;rentgensko slikanje;vratna hrbtenica;radiografska rešetka;doza sevanja;DAP;ED;kakovost slike;

Podatki

Jezik: Slovenski jezik
Leto izida:
Tipologija: 2.09 - Magistrsko delo
Organizacija: UL ZF - Zdravstvena fakulteta
Založnik: [D. Lučić]
UDK: 616-07
COBISS: 234100227 Povezava se bo odprla v novem oknu
Št. ogledov: 68
Št. prenosov: 22
Ocena: 0 (0 glasov)
Metapodatki: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Ostali podatki

Sekundarni jezik: Angleški jezik
Sekundarni naslov: Cervical spine imaging with and without the use of the radiograph grid
Sekundarni povzetek: Introduction: Conventional X-ray imaging of the cervical spine is one of the primary methods for diagnosing cervical spine injuries. The quality of radiographs depends on multiple factors such as the experience of the radiologic technologist, the characteristics of the equipment and imaging techniques. One method to improve image quality is the use of a radiographic grid, which reduces scattered radiation and improves contrast but increases the radiation dose to the patient. Purpose: The aim of the study was to investigate the impact of removing the radiographic grid during lateral cervical spine imaging on the radiation dose and image quality. Methods: The study was conducted in two parts – the first included measurements on a phantom, while the second involved clinical measurements on patients. A descriptive method and a prospective study were employed, with radiographs evaluated by standardized image quality assessment criteria. Radiation dose was measured with a dose area product (DAP) meter. Results: The findings indicate that the removal of the grid led to a statistically significant (p < 0,001) reduction in radiation dose ( 63 % lower dose), without compromising the diagnostic quality of radiographs (p = 0.192). No statistically significant differences (p = 0.226) in the visualization of anatomical structures were identified. The visualization of anatomical structures remained comparable, suggesting that diagnostic image quality can be maintained even without the use of a grid. Discussion and conclusion: Based on these results, gridless imaging presents an effective alternative that optimizes radiological protocols by reducing patient radiation exposure while preserving image quality. Implementing such an approach could significantly contribute to improving patient safety in radiological examinations.
Sekundarne ključne besede: master's theses;radiologic technology;X-ray imaging;cervical spine;radiographic grid;radiation dose;DAP;ED;image quality;
Vrsta dela (COBISS): Magistrsko delo/naloga
Študijski program: 0
Komentar na gradivo: Univ. v Ljubljani, Zdravstvena fak., Oddelek za radiološko tehnologijo
Strani: 66 str., [11] str. pril.