Povzetek
In the article, we emphasize the need for all equipment in a simulation centreto be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. Great care must be exercised when using reduction valves, and pressure testing of the entire circuit is mandatory even if the system is small. We recommend a simple test for gas leakage, performed by placing a finger on possible openings in valves or connecting tubes, which is suitable for use both in a simulation centre and inreal circumstances.
Ključne besede
simulacijski center;oprema;varnost;zapleti;
Podatki
Jezik: |
Slovenski jezik |
Leto izida: |
2011 |
Tipologija: |
1.02 - Pregledni znanstveni članek |
Organizacija: |
UM MF - Medicinska fakulteta |
Založnik: |
Slovensko zdravniško društvo |
UDK: |
617 |
COBISS: |
28569305
|
ISSN: |
1318-0347 |
Matična publikacija: |
Zdravniški vestnik
|
Št. ogledov: |
683 |
Št. prenosov: |
65 |
Ocena: |
0 (0 glasov) |
Metapodatki: |
|
Ostali podatki
Sekundarni jezik: |
Angleški jezik |
Sekundarni naslov: |
Hidden dangers in a simulation centre |
Sekundarni povzetek: |
In the article, we emphasize the need for all equipment in a simulation centreto be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. Great care must be exercised when using reduction valves, and pressure testing of the entire circuit is mandatory even if the system is small. We recommend a simple test for gas leakage, performed by placing a finger on possible openings in valves or connecting tubes, which is suitable for use both in a simulation centre and inreal circumstances. |
Sekundarne ključne besede: |
Educational Technology;Models, Educational;Equipment Safety;Equipment Failure Analysis;Anesthesia;Instrumentation;Nitrogen Oxides;Air;Vacuum;Pulmonary Ventilation;Aparati, odpoved, analiza;Modeli izobraževalni;Anestezija;Aparati, varnost;Izobraževalna tehnologija;Zrak;Dušikovi oksidi;Pljučna ventilacija;Vakuum; |
URN: |
URN:NBN:SI |
Vrsta dela (COBISS): |
Znanstveno delo |
Strani: |
str. 516-522 |
Letnik: |
ǂLetn. ǂ80 |
Zvezek: |
ǂšt. ǂ6 |
Čas izdaje: |
jun. 2011 |
ID: |
9594778 |