Povzetek

In the article, we emphasize the need for all equipment in a simulation centreto be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. Great care must be exercised when using reduction valves, and pressure testing of the entire circuit is mandatory even if the system is small. We recommend a simple test for gas leakage, performed by placing a finger on possible openings in valves or connecting tubes, which is suitable for use both in a simulation centre and inreal circumstances.

Ključne besede

simulacijski center;oprema;varnost;zapleti;

Podatki

Jezik: Slovenski jezik
Leto izida:
Tipologija: 1.02 - Pregledni znanstveni članek
Organizacija: UM MF - Medicinska fakulteta
Založnik: Slovensko zdravniško društvo
UDK: 617
COBISS: 28569305 Povezava se bo odprla v novem oknu
ISSN: 1318-0347
Matična publikacija: Zdravniški vestnik
Št. ogledov: 683
Št. prenosov: 65
Ocena: 0 (0 glasov)
Metapodatki: JSON JSON-RDF JSON-LD TURTLE N-TRIPLES XML RDFA MICRODATA DC-XML DC-RDF RDF

Ostali podatki

Sekundarni jezik: Angleški jezik
Sekundarni naslov: Hidden dangers in a simulation centre
Sekundarni povzetek: In the article, we emphasize the need for all equipment in a simulation centreto be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. Great care must be exercised when using reduction valves, and pressure testing of the entire circuit is mandatory even if the system is small. We recommend a simple test for gas leakage, performed by placing a finger on possible openings in valves or connecting tubes, which is suitable for use both in a simulation centre and inreal circumstances.
Sekundarne ključne besede: Educational Technology;Models, Educational;Equipment Safety;Equipment Failure Analysis;Anesthesia;Instrumentation;Nitrogen Oxides;Air;Vacuum;Pulmonary Ventilation;Aparati, odpoved, analiza;Modeli izobraževalni;Anestezija;Aparati, varnost;Izobraževalna tehnologija;Zrak;Dušikovi oksidi;Pljučna ventilacija;Vakuum;