diplomska naloga
Povzetek
Za določitev nitritov v različnih vzorcih vod (pitnih, podzemnih, površinskih in odpadnih vod) smo uporabili konstantno pretočno analizo. Postopek temelji na izmerjeni absorbanci. Za pripravo osnovne standardne raztopine smo uporabili trden natrijev nitrit.
Rezultate vzorcev smo ovrednotili s pomočjo umeritvene krivulje. Statistično ovrednotenje smo izvedli glede na zahteve analizne metode (koncentracijsko območje, meja določanja, meje zaznavanja, ponovljvost, obnovljivost in pravilnost). S sodelovanjem v medlaboratorijskih primerjavah smo dokazali, da je analizna metoda pravilna.
Rezultate smo ovrednotili glede na dovoljene mejne vrednosti nitritov v pitnih, podzemnih in odpadnih vodah v Sloveniji. Iz rezultatov vidimo, da je vsebnost nitritov v večini analiziranih vzorciev vod nizka, saj ne presega mejnih normativov.
Ključne besede
konstantno pretočni analizator;nitrit;pitne vode;površinske vode;podzemne vode;odpadne vode;validacija;
Podatki
Jezik: |
Slovenski jezik |
Leto izida: |
2009 |
Izvor: |
Maribor |
Tipologija: |
2.11 - Diplomsko delo |
Organizacija: |
UM FKKT - Fakulteta za kemijo in kemijsko tehnologijo |
Založnik: |
[B. Lasecky] |
UDK: |
628.3:543(043.2) |
COBISS: |
13339158
|
Št. ogledov: |
2299 |
Št. prenosov: |
225 |
Ocena: |
0 (0 glasov) |
Metapodatki: |
|
Ostali podatki
Sekundarni jezik: |
Angleški jezik |
Sekundarni naslov: |
Determination of nitrite in waters by continuous flow analysis |
Sekundarni povzetek: |
For determination of nitrite in different water samples like drinking, ground and waste waters the Continuous Flow Analysis (CFA) is used. The analytical procedure is based on measuring absorbance. Stock standard solutions are prepared from sodium nitrite salt. Results of nitrite content in different water samples were evaluated using calibration curve. Statistical evaluation was performed according to the demands of the analytical method (working range, linearity, limit of quantitation LOQ, limit of detection LOD, repeatability, reproducibility and accuracy).
Accuracy of the analytical method has been proven by the participation in the proficiency tests (PT-test).
The obtained results were also evaluated according to the permitted concentration of nitrite in drinking, ground and waste waters in Slovenia. It was established that the concentration of nitrite was below the permitted limit in majority of analysed waters.
|
Sekundarne ključne besede: |
continuous flow analysis (CFA);nitrite;drinking waters;waste waters;validation; |
URN: |
URN:SI:UM: |
Vrsta dela (COBISS): |
Diplomsko delo |
Komentar na gradivo: |
Univ. v Mariboru, Fak. za kemijo in kemijsko tehnologijo |
Strani: |
XII, 56 f., [12] f. pril. |
Ključne besede (UDK): |
applied sciences;medicine;technology;uporabne znanosti;medicina;tehnika;engineering;technology in general;inženirstvo;tehnologija na splošno;public health engineering;water;sanitation;illuminating engineering;sanitarna tehnika;zdravstvena tehnika;voda;sanitarne naprave;svetlobna tehnika;sewage;treatment;disposal;utilization of sewage;odpadne vode;mathematics;natural sciences;naravoslovne vede;matematika;chemistry;crystallography;mineralogy;kemija;analytical chemistry;analizna kemija; |
ID: |
986769 |