doktorska disertacija
Povzetek
Spremljanje aktivnosti srca v obliki frekvence srčnega utripa (angl.: heart rate, HR) je ena izmed
najbolj rutinskih meritev, ki se izvaja v sklopu ugotavljanja osnovnega zdravstvenega stanja
posameznika. Povprečen HR zdrave odrasle osebe v mirovanju je med 60 in 100 utripi na minuto,
pri športno aktivnih ljudeh pa lahko tudi manj. Pomembno odstopanje od povprečnih vrednosti
nakazuje možnost bolezni srca in ožilja. Poleg HR se je v medicini v zadnjih letih kot kvantitativen
pokazatelj napovedovanja verjetnosti obolenj na področju srca in ožilja uveljavil tudi parameter
variabilnosti frekvence srčnega utripa (angl.: heart rate variability, HRV). Slednjega dobimo na
podlagi matematične analize HR v časovni ali frekvenčni domeni. Zaradi navidezno lahkega
izračuna pripadajočih parametrov se je uporaba HRV razširila. Kljub temu sta pomembnost in
pomen posameznih parametrov bolj kompleksna od splošnega prepričanja. Morebitni napačni
izračuni ali njihova interpretacija lahko vodijo v napačne zaključke in pretirane ekstrapolacije. Za
pravilen izračun HRV je pomembna zanesljivost merilne metode za pridobivanje HR.
Najpogosteje se HR in HRV merita v kliničnih aplikacijah. Zaradi svoje zanesljivosti in relativno
enostavnega izvajanja veljata za standardni metodi za izvajanje meritev elektrokardiograf (EKG)
ter optična pletizmografija (PPG). Obe metodi se izvajata v kontaktnem načinu, zaradi česar sta
manj primerni za določene skupine bolnikov, kot so na primer bolniki z opeklinami, nedonošenčki,
itd. Poleg tega se v medicini in tudi na pretežno gospodarskih področjih pojavlja vedno več potreb
po dolgotrajnem izvajanju meritev, tudi na delovnih mestih. Omenjeni standardni metodi zaradi
senzorjev, ki morajo biti v kontaktu z merjencem, ovirata opravljanje vsakdanjih ali poklicnih
aktivnosti, poleg tega pa morebitni premiki senzorjev zmanjšajo razmerje signal – šum. Zaradi
tega se je pojavila potreba po brezkontaktni izvedbi meritev. Cilj doktorske disertacije je
predstavitev nove brezkontaktne metode za merjenje HR in HRV. Doktorska disertacija je
usmerjena izključno v merilno problematiko v inženirskem in znanstveno-raziskovalnem smislu.
Medicinski vidiki so upoštevani in privzeti s strani strokovnjakov s področja medicinske znanosti.
V tem doktorskem delu opisujem predlog brezkontaktnega ultrazvočnega merjenja srčnih
parametrov HR in HRV. Metoda temelji na Dopplerjevem pojavu, kjer merimo razliko v fazi med
oddanim in odbitim signalom na izbranem področju telesa, npr. na vratu. Do razlike v fazi med
signaloma pride zaradi fizičnega premika kože, kot posledice utripanja površinske karotidne
arterije. Novost v primerjavi z ostalimi študijami je sočasna uporaba ultrazvočnega signala na dveh
frekvencah. Z možnostjo izbire optimalnega signala v realnem času pri eni ali drugi frekvenci smo
znatno izboljšali zanesljivost metode in točnost rezultatov zaradi izničenja stojnega vala in drugih
morebitnih okoliških motenj. Signal iz senzorja smo zajemali z A/D pretvornikom, vzbujevalni
signal pa generirali z D/A pretvornikom. Obdelava signala je sledila v namensko razviti aplikaciji
v okolju LabVIEW, ki je poleg spremljanja meritev v realnem času omogočala tudi snemanje
signala za potrebe kasnejše dodatne obdelave. Eksperimentalno metodo smo neposredno
primerjali z referenčno kontaktno EKG metodo preko sistema Biopac MP150 z modulom
ECG100C.
Predlagana metoda je bila preizkušena v treh fazah. V prvi fazi smo simulirali utripanje žile z
merjenjem premikanja membrane zvočnika, priključenega na signalni generator. Najprej smo
merili signal pri konstantni frekvenci, nato pa smo frekvenco v poljubnem časovnem intervalu
poviševali. V drugi fazi smo izvedli meritev na prostovoljcih, ki so ležali na postelji v laboratorijskem okolju. Izvedena je bila v dveh korakih pri različnih HR, in sicer pri nižjem med
počitkom, ter pri povišanem po 1 minuti fizične aktivnosti. Zadnji del študije smo izvedli v realnem
kliničnem okolju. Na Oddelku za kardiologijo v Univerzitetnem kliničnem centru v Ljubljani
(UKCLJ) smo brezkontaktno merilno metodo uporabili na skupini prostovoljcev z različnimi
boleznimi srca in ožilja. Za vse meritve v laboratorijskem in kliničnem okolju smo izračunali
statistične parametre HRV v časovni domeni v primerjavi z referenčnim signalom.
Pred izračunom statističnih parametrov smo iz posnetega signala odstranili dele signala z znano
motnjo (npr. premikanje posameznika, požiranje sline, itd.). Izračunali smo srednjo vrednost in
standardni odklon razlik med trenutnim HR eksperimentalnega signala pri optimalni frekvenci v
primerjavi z referenčnim signalom, ter srednjo vrednost in standardni odklon razlik med trenutnim
HR eksperimentalnega signala, ki smo ga primerjali z premičnim povprečjem, neodvisno od
referenčnega EKG signala. V prvem primeru smo za skupino prostovoljcev v laboratorijskem
okolju med fazo mirovanja dobili ΔHR = 0,23 min-1 ± 0,61 min-1, v drugem pa ΔHR = 0,31 min-1
± 0,88 min-1. Po fizični aktivnosti smo dobili v prvem primeru ΔHR = 0,29 min-1 ± 0,67 min-1, v
drugem pa ΔHR = 0,42 min-1 ± 0,75 min-1. Rezultat za primer kliničnega preizkusa, ki smo ga
podali v delu, je ΔHR = -0,41 min-1 ± 1,97 min-1 pri primerjanju z referenčno metodo, ter ΔHR =
-0,50 min-1 ± 2,00 min-1 pri premičnem povprečju. Za HRV parametre v časovni domeni smo
izračunali relativni pogrešek med referenčnim in eksperimentalnim signalom. Za skupino znotraj
laboratorijskega dela so rezultati meritev med mirovanjem AVNN = 0,07 % ± 0,13 %, SDNN =
1,92 % ± 3,54 %, rMSSD = 9,05 % ± 13,93 %, pNN20 = 5,16 % ± 3,03 %, pNN50 = 3,09 % ±
3,53 % pri signalu dobljenem s primerjavo z EKG. Po fizični aktivnosti pa AVNN = 0,06 % ± 0,05
%, SDNN = 0,88 % ± 0,48 %, rMSSD = 11,44 % ± 16,66 %, pNN20 = 25,09 % ± 30,86 %, pNN50
= 1,54 % ± 0,86 %. Za podan primer meritev v kliničnem eksperimentu pa so HRV parametri
sledeči AVNN = 0,05 %, SDNN = 32,82 %, rMSSD = 4,04 %, pNN20 = 8,94 %, pNN50 = 6,32
%.
Pilotske raziskave na skupini prostovoljcev v laboratorijskem in tudi v kliničnem okolju kažejo na
to, da je brezkontaktno merjenje fizioloških parametrov s to metodo možno. Rezultati naše študije
kažejo tudi, da je metoda ponovljiva in primerljiva s standardnimi kontaktnimi merjenji.
Ključne besede
frekvenca srčnega utripa;spremenljivost frekvence srčnega utripa;brezkontaktno merjenje;ultrazvok;radar;študije izvedljivosti;klinično ovrednotenje;disertacije;
Podatki
Jezik: |
Slovenski jezik |
Leto izida: |
2017 |
Tipologija: |
2.08 - Doktorska disertacija |
Organizacija: |
UL FE - Fakulteta za elektrotehniko |
Založnik: |
[J. Kranjec] |
UDK: |
621.317:612.16(043.3) |
COBISS: |
11924820
|
Št. ogledov: |
2370 |
Št. prenosov: |
835 |
Ocena: |
0 (0 glasov) |
Metapodatki: |
|
Ostali podatki
Sekundarni jezik: |
Angleški jezik |
Sekundarni naslov: |
NEW METHODS OF NON-CONTACT MEASURING THE HEART RATE VARIABILITY |
Sekundarni povzetek: |
The monitoring of heart activity in the form of heart rate (HR) is one of the most routine
measurements conducted in order to determine the basic medical condition of an individual. The
standard average HR of a healthy adult individual is between 60 and 100 beats per minute and
could be lower for individuals in good physical condition. A noticeable deviation from these
standard values indicates a possibility of cardiovascular diseases. In the last couple of decades
heart rate variability (HRV) has been recognized as an important quantitative marker of the
relationship between the autonomic nervous system and cardiovascular mortality, including
sudden cardiac death. The apparent easy derivation by performing mathematical analysis from the
HR has popularized its use. However, the significance and meaning of many different measures of
HRV are more complex than generally appreciated and there is a potential for incorrect conclusions
and for excessive extrapolations. Because of this, the reliability of HR and HRV measuring device
is of utmost importance.
HR and HRV are most frequently measured and observed in clinical applications. Because of their
reliability and easy implementation, electrocardiography (ECG) and optical plethysmography
(PPG) are considered as the “gold standard”. Both are performed in a contact manner, which is
why they are less appropriate for specific groups of patients, e.g. patients with burns, neonates,
etc. Next to this, there is more and more interest of continuous and long-term measurements also
in predominantly economic areas. The mentioned standard methods operate with sensors, which
need to be in direct contact with human skin and therefore limit the everyday or occupational
activities. Furthermore, any possible unwanted movements of the sensors’ location on the skin
may result in noise or other signal deformation, which could make it inappropriate for analysis. As
a result, the need for non-contact measurement process has emerged. The goal of this dissertation
is to present a new non-contact method for measuring of the HR and the HRV. The dissertation is
focused exclusively on the measurement problematics in the engineering and scientific research
sense. Medical aspects are taken into account, provided by the experts in the field of medical
science.
Within this thesis we describe a proposal of non-contact ultrasound measuring of the HR and the
HRV. The method is based on the Doppler phenomenon as it measures the difference in phase
between the transmitted and deducted signal on the selected region of interest, for example on a
person’s neck. The difference of the phase between the two signals is caused by the pulsation of
the carotid artery. The novelty of our method, compared to other studies, is transmitting at two
different ultrasound frequencies at the same time. With the option of selecting the optimal signal
in real time at either frequency, the reliability of the method and the accuracy of the results have
been significantly improved due to the elimination of the standing waves and other possible
environmental disturbances. The measuring device’s signal was captured with an A/D converter
device, whereas the excitation signal was generated with a D/A converter device. Signal processing
was carried out in a dedicated LabVIEW application, which in addition to monitoring in real-time
also enables the recording of a signal for later processing. The results from the experimental
method were directly compared to the reference contact ECG method, obtained with a Biopac
system MP150 with the ECG100C module. The proposed method was tested in three phases. In the first phase, we simulated the contraction
of a vein by measuring the pulsation of a loudspeaker membrane, which was directly connected to
a signal generator. First, we measured the signal at a constant frequency and later the frequency
was raised in a non-specific time interval. In the second phase, we performed measurements on
volunteers who were lying in bed in laboratory environment. This phase was performed by
measuring at different HR. Firstly at lower HR during rest and secondly at elevated HR after 1
minute of exercise. The last phase of the study was carried out in a realistic clinical setting at the
Clinical department of cardiology, University Medical Centre of Ljubljana (UKCLJ). The noncontact
measuring method was performed on a group of patients with various cardiovascular
diseases. For all measurements in the laboratory and clinical environment, HRV time domain
statistical parameters were calculated.
Prior to calculating the statistical parameters, the parts of the signal with known noise artefact were
removed from the recorded signal (for instance individual’s movement, swallowing of saliva, etc.).
The calculated mean value and standard deviation of the differences between the current HR of
the experimental signal at the optimal frequency and the reference signal on the one hand, and on
the other hand the values were obtained based on moving average of the non-contact signal,
independently of the ECG signal. During the standstill phase, in the first case we obtained results
for the group of volunteers in the laboratory environment ΔHR = 0,23 min-1 ± 0,01 min-1, and in
the second case ΔHR = 0,31 min-1 ± 0,88 min-1. After 1 minute of physical activity the following
values were calculated for the first case ΔHR = 0,29 min-1 ± 0,67 min-1 and ΔHR = 0,42 min-1 ±
0,75 min-1 for the second case. The result of the clinical trial described in this dissertation was
ΔHR = -0,41 min-1 ± 1,97 min-1 when compared to the reference method, and ΔHR = 0,50 min-1 ±
2,00 min-1 when calculated based on the moving average. For the HRV parameters in the time
domain, we calculated the relative error between the reference and the experimental signal. For
the group within the laboratory setting, the results of measurements during standstill phase were
AVNN = 0,07 % ± 0,13 %, SDNN = 1,92 % ± 3,54 %, rMSSD = 9,05 % ± 13,93 %, pNN20 = 5,16
% ± 3,03 %, pNN50 = 3,09 % ± 3,53 %. After physical activity, the results were as follows: AVNN
= 0,06 % ± 0,05 %, SDNN = 0,88 % ± 0,48 %, rMSSD = 11,44 % ± 16,66 %, pNN20 = 25,09 % ±
30,86 %, pNN50 = 1,54 % ± 0,86 %. For the given example of measurements in the clinical
experiment, the HRV parameters were AVNN = 0,05 %, SDNN = 32,82 %, rMSSD = 4,04 %,
pNN20 = 8,94 %, pNN50 = 6,32 %.
The pilot research on a group of volunteers in the laboratory and in the clinical setting suggests
that the non-contact measurement of physiological parameters with the proposed method is indeed
possible. Moreover, the results of our study show that the proposed non-contact measurement
method is reproducible and comparable to standard contact measurement methods. |
Sekundarne ključne besede: |
Srčna frekvenca;Disertacije;Merjenje; |
Vrsta dela (COBISS): |
Doktorsko delo/naloga |
Študijski program: |
1000319 |
Konec prepovedi (OpenAIRE): |
1970-01-01 |
Komentar na gradivo: |
Univ. v Ljubljani, Fak. za elektrotehniko |
Strani: |
XII, 120 str. |
ID: |
10915151 |